
CORN PROCESSING
Probe-Sampling of Grain Corn and the Code of Practice
By Brian Doidge, Ridgetown College, University of Guelph
In March, the Canadian Grain Commission (CGC) submitted the final report for
a project the OCPA had commissioned to assess whether probe-sampling procedures used at processor locations affected
the grading of corn by influencing the Cracked Corn and Foreign Material percentage (CCFM) of corn samples. The
CGC conducted on-site comparisons of probe-sampling results at King Milling, Chatham; Commercial Alcohols, Chatham;
and Casco, London. The OCPA thanks these processing plants for their cooperation in the project.
At each site, 30 trucks of grain corn were probed three ways: in the processor’s usual probe-sampling procedure;
by the processor using the CGC’s probe-sampling procedure; and then hand sampled by the CGC. Hand sampling is considered
by the CGC to be the standard for comparison. All samples were graded by the CGC and results from the three sampling
procedures compared.
For all locations, hand sampling gave an average CCFM percentage 0.7 per cent lower than probe-sampling using the
CGC probe procedure, and 0.8 per cent lower than probe-sampling using the processor’s probe procedure. Remember,
this is the same corn (1997 crop corn, which was of exceptional quality, sampled in the late summer of 1998) from
the same truck using the same probe equipment and personnel...only the probe procedure changed.
CCFM is a grading factor for corn under the Canada Grain Act Official Grain Grading Guide:
- Grade No. 1 CW/CE 2.0% CCFM
- Grade No. 2 CW/CE 3.0% CCFM
- Grade No. 3 CW/CE 5.0% CCFM
- Grade No. 4 CW/CE 7.0% CCFM
- Grade No. 5 CW/CE 12.0% CCFM
Therefore, for the 90 loads of corn sampled (30 at each of three processor locations),
because of the CCFM result:
- The processor probe procedure had 20 (22.2 per cent) that were lower grades than
for hand-sampling.
- The CGC probe procedure had 14 (15.5 per cent) that were lower grades than for
hand-sampling.
- Hand-sampling had three (3.3 per cent) that were lower grades than for the processor
probe procedure.
- Hand-sampling had two (2.2 per cent) that were lower grades than for the CGC
probe procedure.
Because none of these processors can utilize (and therefore, do not accept) corn
lower than Grade No. 3, and because of the large discount applied against Grade No. 3 corn versus Grade No. 2 corn
at processor locations (i.e., $3 - $5/mt in many instances), down-grading from No. 2 to No. 3 on account of CCFM
content is of most concern.
Of the 20 samples (22.2 per cent of the 90 samples) that were lower graded by processor probe procedure versus
hand-sampling:
- 13 moved from a No. 2 down to a No. 3.
- Two moved from a No. 2 down to a No. 4.
- Four moved from a No. 3 down to a No. 4.
- One moved from a No. 3 down to a No. 5.
Of the 14 samples (15.5 per cent of the 90 samples) that were lower graded by
CGC probe procedure versus hand-sampling:
- 11 moved from a No. 2 down to a No. 3.
- Two moved from a No. 3 down to a No. 4.
- One moved from a No. 3 down to a No. 5.
A project steering committee reviewing these and other
CGC results suggests the following:
- Spreaders on loading spouts at the farm would spread the corn and fines more
evenly throughout the load, rather than concentrating fines in the centre of the load when loaded without benefit
of a spreader. Even distribution of fines would improve samples and make them more representative of the load.
- The study suggests hopper-bottom trucks and trailers provide better quality and
more representative samples compared to flat-bottom trucks and trailers.
- Probe design, age and set-up affects breakage, depending on distance from probe
to scale-house, the number of bends and material of the sample transport piping.
- Detailed inspection of probe equipment should be conducted several times during
the season with attention on efficiency of probe door sealment, to prevent vacuum leakage when the sample is pneumatically
transported from the load to the scale-house.
- The more probe samples taken per load the better, and the closer the result will
be to hand-sampling.
- The higher the ratio of the number of probes on the outside of the load compared
to the number of probes in the centre of the load, the more representative will be the sample. For example, four
probes taken from the outside (i.e., four corners) to every one probe from the centre will provide a more representative
sample than only twoprobes from the outside to every one probe from the centre (i.e., a diagonal across the truck).
Recommended procedure in the U.S. is a ratio of 6:1 or 8:1.
- The probe must go straight down into the load, not on an angle toward the centre
of the load.
- Hand sampling to calibrate probe-sample results perhaps should be conducted randomly
(daily?) during the entire season to verify probe sample results and to signal to shippers that loads are indeed
being inspected closely.
- Processor grading personnel should be trained by CGC personnel on proper techniques
and equipment for hand-sampling.
- Probe-sample CCFM results might be adjusted frequently throughout the season
based on randomized hand-sampling calibration results to compensate for changes in the corn profile being received.
- Probe sampling CCFM ratings might better be adjusted as a percentage of the probe
CCFM result rather than an absolute number. For example, 0.2 per cent lower at a probe CCFM reading of 2.2 per
cent; 0.7 per cent lower at a probe CCFM reading of 3.4 per cent, etc.
Processors have responded to the CGC project and steering committee suggestions
with these actions:
- King Milling, even before these results were in, had conducted its own comparison
testing of probe sampling versus hand-sampling, had compared results and created its own “line of best fit” (i.e.,
adjustments necessary at various probe CCFM readings in order to align with hand-sampling readings), and then applied
this adjustment as a standard “correction factor” to CCFM results.
- Casco “has taken the approach that we are willing to take additional probes from
the outside of the truck in order to parallel hand-sampling. Casco feels that if they take one sample from the
centre of the truck and three to four from the side of the truck this should help eliminate the discrepancies for
CCFM.”
- Commercial Alcohols’ corn receiving and grading criteria are similar to Casco’s,
as Casco purchases all corn for the ethanol plant in Chatham. Results of this study put both producers and processors
in an awkward situation concerning sampling of grain corn at processor delivery locations. Producers should know
that these processors have agreed to abide by the Code of Practice for Trade in Corn & Canola in Ontario, which
governs deliveries at primary elevator locations in the province. The code states that “the elevator’s representative
receiving the corn or canola, or the producer, or the producer’s agent, may require a sample of corn or canola
to be taken. Such sample shall be accepted by all parties as a fair and representative sample of the corn or canola
from which it was taken.” The code makes it clear that the producer and the processor must agree that the sample
taken from the load is a fair and representative sample. So, when grain corn grading depends on samples gathered
by probes, before accepting a probe sample as representative of a load suitable for testing (especially if the
grading result is or might be in dispute) producers should note carefully:
- that the probe went into the load straight down, not angled into the centre
- that there were four probe samples taken from the outside versus 1 from the center
- that probe sampling results in a CCFM percentage about 0.75 per cent higher than
hand-sampling, even if both the first two criteria are present. At the request of the producer or the producer’s
representative (this would allow a trucker to request), or the representative of the grain receiver, a representative
sample of the grain corn shall be taken by probe-sampling procedure as outlined, or by hand-sampling using accepted
CGC practice.
It is suggested that prior to resorting to arbitration of disputes concerning
grading on account of CCFM content, producers and processors negotiate acceptable resolutions, considering the
results of the March 1999 CGC Probe-Sampling project.
1